Determination of sodium artesunate in plasma using ion-pairing high-performance liquid chromatography

Autor: R.G. Reid, R.R. Moody, M.I Awad, R.B. Taylor
Rok vydání: 2000
Předmět:
Zdroj: Journal of chromatography. B, Biomedical sciences and applications. 744(2)
ISSN: 1387-2273
Popis: A chromatographic method is described for the determination of sodium artesunate in plasma. This includes cetyltrimethylammonium bromide as a cationic pairing ion in a reversed-phase system using an octadecylsilica 100×4.6 mm I.D. 3 μm analytical column with a mobile phase of acetonitrile/acetate buffer at pH7. Column switching incorporating a 5 μm octadecylsilica 100×4.6 mm I.D. precolumn is used in addition to off-line solid-phase extraction for pretreatment of plasma samples in order to eliminate interference from endogenous components. Detection is by post-column derivatisation with 1.0 M methanolic KOH followed by UV detection at 289 nm. Calibration is linear over the range 100–1600 ng ml −1 and the limit of detection is estimated as 20 ng ml −1 . Illustrative results are shown of the artesunate plasma levels determined by the proposed method following the administration of artesunate as tablets and as suppositories to healthy volunteers.
Databáze: OpenAIRE