Strategy to Eliminate Inappropriate Shocks Secondary to T-Wave Oversensing in a Biventricular ICD
Autor: | Jonathan S. Silver, Mary Ellen Gray, Roy M. John |
---|---|
Rok vydání: | 2009 |
Předmět: |
medicine.medical_specialty
business.industry Heart Ventricles General Medicine Defibrillators Implantable Electric Injuries Electrocardiography Internal medicine medicine Cardiology Humans Equipment Failure Female In patient Cardiology and Cardiovascular Medicine Intensive care medicine business Aged |
Zdroj: | Pacing and Clinical Electrophysiology. 32:134-136 |
ISSN: | 1540-8159 0147-8389 |
DOI: | 10.1111/j.1540-8159.2009.02187.x |
Popis: | T-wave oversensing represents a common cause of inappropriate shocks in patients with implanted cardiac defibrillators. This case report demonstrates a strategy of device programming using V-V pace delay (sequential rather than simultaneous biventricular pacing) to eliminate T-wave oversensing without decreasing sensitivity to detect true tachyarrhythmia. |
Databáze: | OpenAIRE |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |