Micro-computed tomography evaluation of general trends in aligner thickness and gap width after thermoforming procedures involving six commercial clear aligners: An in vitro study

Autor: Mattia Longo, Mario Palone, Giorgio Alfredo Spedicato, Giuseppe Siciliani, Fabio De Pascalis, Luca Lombardo, Michele Nacucchi, Niki Arveda
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Korean Journal of Orthodontics
Popis: Objective To assess the effects of thermoforming on aligner thickness and gap width in six aligner systems with the same nominal thickness. Methods Six passive upper aligners of different brands were adapted to a single printed cast. Each sample was evaluated with high-resolution micro-computed tomography. To investigate aligner thickness and gap width, two-dimensional (2D) analysis was conducted assessing the effects of the following variables: tooth type (central incisor, canine, and first molar), 2D reference points, and aligner type. Data were analyzed and compared using analysis of variance and Tukey's post-hoc tests (p < 0.05). Results Tooth type, dental region, and aligner type affected both the gap width and aligner thickness. The aligner thickness remained moderately stable across the arch only in the F22. Conclusions All thermoformed samples displayed smaller aligner thickness and gap width at anterior teeth and both gingival and coronal centers than at posterior teeth and occlusal surfaces.
Databáze: OpenAIRE