Bone ingrowth into open architecture PEEK interference screw after ACL reconstruction

Autor: Bjarne Mygind-Klavsen, Torsten Toftegaard Nielsen, Peter Faunø, Ole Gade Sørensen, Martin Lind, Stacy Leake-Gardner
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Journal of Experimental Orthopaedics, Vol 7, Iss 1, Pp 1-8 (2020)
Journal of Experimental Orthopaedics
Lind, M, Nielsen, T G, Sørensen, O G, Mygind-Klavsen, B, Faunø, P & Leake-Gardner, S 2020, ' Bone ingrowth into open architecture PEEK interference screw after ACL reconstruction ', Journal of experimental orthopaedics, vol. 7, no. 1, 68 . https://doi.org/10.1186/s40634-020-00285-z, https://doi.org/10.1186/s40634-020-00285-z
ISSN: 2197-1153
Popis: Purpose Open or fenestrated interference screw design that allow bone ingrowth is a concept for improved bone healing to softtissue graft and bone filling in bone tunnels after anterior cruciate ligament reconstruction (ACLR) The aim of the current study was to assess CT scanning evaluated bone ingrowth into an open architecture interference screws in the tibial tunnel of patients undergoing ACL with soft tissue grafts. It was hypothesized that open architecture interference screws would stimulate bone ingrowth into the screw cavities. Methods Twelve patients requiring ACLR were included. They underwent arthroscopic ACLR with semitendinosus−/gracilis tendon graft and an open architecture polyetheretherketone (PEEK) interference screw. The patients were scanned with a multi-slice CT scanner two weeks, six and twelve months postoperatively. On CT reconstruction slices bone ingrowth into the screw was measured. Subjective and objective clinical outcome international knee documentation committee score and instrumented knee laxity determination were collected. Results At six months no implants demonstrated more than 10% bone ingrowth. At twelve months 42% (5/12) implants had more than 10% bone ingrowth (p = 0.009). The average bone filling into the screws was 7.7%. There was no tunnel widening or cyst formation seen in relation to any of the implants. Subjective IKDC score improved significantly from 50.6 baseline to 80.1 at 24 month follow-up. Preoperative side-to-side knee laxity improved from 3.7 (2.1) to 1.4 (1.2) mm at twelve months. There were no serious adverse events in relation to the new open architecture thread PEEK interference screw during or after hamstring ACL reconstruction. Conclusion The present study demonstrated that open architecture thread PEEK interference screw can stimulate bone ingrowth into the screws after soft tissue ACL reconstruction with at 12 months with an average bone filling into screws was 7.7%. Knee stability, functional, subjective and objective outcomes were similar to large volume ACL outcome studies. Trial registration The study was registered at ClinicalTrials # NCT02382341. 12-09-2014. Level of evidence IV.
Databáze: OpenAIRE