Limitations of the zero crossing detector in the analysis of intracoronary doppler: A comparison with fast fourier transform analysis of basal, hyperemic, and transstenotic blood flow velocity measurements in patients with coronary artery disease
Autor: | Peter de Jaegere, Patrick W. Serruys, David T. Linker, J.A.F. Oomen, Carlo Di Mario, Jos R.T.C. Roelandt |
---|---|
Rok vydání: | 1993 |
Předmět: |
Male
medicine.medical_specialty Spectrum analyzer medicine.medical_treatment Transducers Fast Fourier transform Coronary Disease Coronary circulation symbols.namesake Coronary Circulation Papaverine Angioplasty medicine Humans Ultrasonics Angioplasty Balloon Coronary Ultrasonography Fourier Analysis business.industry Coronary flow reserve Signal Processing Computer-Assisted Blood flow Middle Aged Coronary Vessels Surgery medicine.anatomical_structure Flow velocity symbols Regression Analysis Female Cardiology and Cardiovascular Medicine business Nuclear medicine Doppler effect Algorithms Blood Flow Velocity |
Zdroj: | Catheterization and Cardiovascular Diagnosis. 28:56-64 |
ISSN: | 1097-0304 0098-6569 |
DOI: | 10.1002/ccd.1810280112 |
Popis: | The current clinical standard for the analysis of intracoronary Doppler signals is the application of a zero-crossing (ZC) detector. However, the accuracy of the method is questionable, especially in areas of disturbed flow, as confirmed by in vitro studies, animal experiments, and intraoperative observations. The aim of this study is the comparison of a conventional ZC detector and a custom-designed spectral analyzer (fast Fourier transform, FFT) in the analysis of intracoronary Doppler signals obtained in 19 patients undergoing coronary angioplasty. A 3F catheter with an end-mounted Doppler ceramic crystal was placed over an 0.014" guidewire in a normal or near-normal segment proximal to the lesion to be dilated. The Doppler signal was recorded before and after intracoronary infusion of 12.5 mg of papaverine. In 9 patients high flow velocities could be recorded when the catheter was advanced across the stenosis. The blood flow velocity measurements obtained with ZC were significantly lower than the maximal FFT flow velocity measurements (16 +/- 12 cm/s vs. 29 +/- 18 cm/s, p < .001). In all the conditions of Doppler signal acquisition (baseline, hyperemia, stenosis) a large scattering of the signed differences between corresponding measurements was observed. The standard deviation of the difference ZC-FFT was +/- 11 cm/s and +/- 5 cm/s for the maximal and mean FFT flow velocity, corresponding in both cases to +/- 37% of the mean of the ZC and FFT measurements.(ABSTRACT TRUNCATED AT 250 WORDS) |
Databáze: | OpenAIRE |
Externí odkaz: |