Energy expenditure measurements are reproducible in different whole‐room indirect calorimeters in humans
Autor: | Emma J. Stinson, Theresa Rodzevik, Jonathan Krakoff, Paolo Piaggi, Douglas C. Chang |
---|---|
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | Obesity. 30:1766-1777 |
ISSN: | 1930-739X 1930-7381 |
Popis: | This study aimed to evaluate the agreement of commonly reported energy metabolism measurements obtained from two different whole-room indirect calorimeters (WRICs).Nine healthy adult volunteers were evaluated over four separate 24-hour periods in a crossover design, twice in two different WRICs of different sizes, each operated according to the Room Indirect Calorimetry Operating and Reporting Standards published in 2020. The reproducibility of repeated measurements was quantified by the coefficient of variation (CV) and intraclass correlation coefficient (ICC).The CVs between and within each WRIC for average 24-hour carbon dioxide production rate (VCOCross-chamber reproducibility is high for common outcome measures assessed in the respiratory chamber. The results support efforts to promote standardization across WRICs to allow multicenter studies. |