Comparison of Raman Spectrograph Throughput Using Two Commercial Systems: Transmissive versus Reflective
Autor: | Elizabeth M. Kanter, Anita Mahadevan-Jansen, Chad A. Lieber |
---|---|
Rok vydání: | 2008 |
Předmět: |
Materials science
Spectrophotometry Infrared Spectrometer business.industry Instrumentation Resolution (electron density) Equipment Design Spectrum Analysis Raman Equipment Failure Analysis symbols.namesake Optics symbols Fiber Optic Technology Optoelectronics Spectrophotometry Ultraviolet Photonics business Raman spectroscopy Throughput (business) Spectrograph Spectroscopy Raman scattering |
Zdroj: | Applied Spectroscopy. 62:575-582 |
ISSN: | 1943-3530 0003-7028 |
Popis: | A common goal of most Raman spectroscopists is the acquisition of clear Raman spectra in short integration times. It is paramount, then, that the hardware used in a Raman system be optimized for throughput, signal-to-noise ratio, and resolution. To this end, we performed a systematic comparison of two Raman spectrographs: the widely used Kaiser HoloSpec™ and a relatively new commercial offering, the Raman Explorer™ by Headwall Photonics. Both strong and weak Raman scattering samples were measured using various launching conditions. When resolution-matched, the throughputs of both spectrographs were found to be roughly similar over the central range of the fingerprint region (approximately 800 to 1140 cm−1, using 785 nm illumination), with the Raman Explorer™ demonstrating a slight throughput advantage outside this range. Other factors are also considered such that end users may better select the optimum spectrograph for their particular application. |