Measurement of regional cerebral blood flow. A comparison of the mass spectrographic and intra-arterial xenon 133 techniques
Autor: | Melville Roberts, David J. Silver, Guy Owens, Robert Rembish |
---|---|
Rok vydání: | 1974 |
Předmět: |
Xenon
Partial Pressure chemistry.chemical_element Mass spectrometry pCO2 Mass Spectrometry Intra arterial Methods Animals Argon business.industry Computers Washout Blood flow respiratory system Carbon Dioxide Cannula respiratory tract diseases Oxygen Cerebral blood flow chemistry Evaluation Studies as Topic Cerebrovascular Circulation Cats Neurology (clinical) Nuclear medicine business circulatory and respiratory physiology |
Zdroj: | Neurology. 24(4) |
ISSN: | 0028-3878 |
Popis: | Regional cerebral blood flow in the cat was quantitatively measured by Xe133 clearance and mass spectrographic monitoring of argon washout. No statistical difference was observed in the values obtained for regional cerebral blood flow by the two methods. Although the mass spectrometer cannula caused some brain damage, since brain PO2 and pCO2 were monitored simultaneously, significant tissue trauma was detected early. The mass spectrographic technique enables precise measurement of blood flow in any locality, and tissue PO2 and pCO2 can be simultaneously recorded. |