Prevalence of Complications During Implantation and During 38-Month Follow-up of 1060 Consecutive Patients With Implantable Cardioverter-Defibrillators

Autor: Fausan S. Tsai, Adam Kruger, Wilbert S. Aronow, Hoang M. Lai, Martin H. Cohen, Harit Desai, Srikala Devabhaktuni, Carmine Sorbera, William H. Frishman
Rok vydání: 2010
Předmět:
Zdroj: American Journal of Therapeutics. 17:e8-e10
ISSN: 1075-2765
DOI: 10.1097/mjt.0b013e3181855c6b
Popis: During implantation and during 38-month follow-up of 1060 consecutive patients who had implantable cardioverter-defibrillators, complications occurred in 60 (5.7%) of 1060 patients. These complications consisted of fractured leads requiring lead revision in 36 (3.4%) patients, lead infection requiring antibiotics in 5 (0.5%) patients, device replacement because of malfunction in 5 (0.5%) patients, repositioning of leads in 3 (0.3%) patients, a hematoma at the time of implantation in 3 (0.3%) patients, pneumothorax at the time of implantation in 2 (0.2%) patients, repair of a defective generator in 1 (0.1%) patient, replacement of the device because of atrophy of the skin over the device in 1 (0.1%) patient, a transient ischemic attack because of atrial fibrillation developing during implantation in 1 (0.1%) patient, device replacement because of a recall from Guidant in 1 (0.1%) patient, pocket revision because of pain when lying on the side of the pacemaker in 1 (0.1%) patient, and pacemaker infection in 1 (0.1%) patient.