Oxygen therapy via high-flow nasal cannula generates physiologic changes that impact ventilatory variables of patients. However, we know that there are detrimental effects on airway mucosa related to inhalation of gases. The objective of this study was to evaluate the performance in terms of absolute humidity, relative humidity, and temperature of different brands of heated humidifiers and circuits in the invasive mode during the use of high-flow oxygen therapy in flows between 30 and 100 L/min.A prospective observational study conducted at the Sanatorio Anchorena equipment analysis laboratory; September 5 to October 20, 2019.A statistically significant interaction was found among the programmed flows and the different combinations of devices and circuits for the delivery of absolute humidity (When heated humidifiers were used in the invasive mode for high-flow oxygen therapy, absolute humidity depended not only on the heated humidifiers and the combination of circuits but also on the programmed flow, especially at flows50 L/min. Moreover, the heated humidifiers exhibited different behaviors, in some cases inefficient, in delivering adequate humidification. However, some equipment improved performance when set to the invasive mode. |