Dedicated plug based closure for large bore access –The MARVEL prospective registry
Autor: | Janarthanan Sathananthan, Mika Laine, Stefan Toggweiler, Jur ten Berg, Noriaki Moriyama, Federico Moccetti, Christian Juhl Terkelsen, Mikko Savontaus, Ton Slagboom, Peter P de Jaegere, Herbert Kroon, Evald Høj Christiansen, Nicolas M. Van Mieghem, David A. Wood, Giovanni Amoroso, Livia L. Gheorghe, John G. Webb, Guus R. G. Brueren, Pim A.L. Tonino, Markus Malmberg, Joost Daemen |
---|---|
Přispěvatelé: | HUS Heart and Lung Center, Kardiologian yksikkö, Helsinki University Hospital Area, Department of Medicine, Cardiology |
Jazyk: | angličtina |
Rok vydání: | 2020 |
Předmět: |
vascular complications
Percutaneous DEVICES IMPACT medicine.medical_treatment 030204 cardiovascular system & hematology Balloon Endovascular aneurysm repair DISEASE Original Studies 0302 clinical medicine Valve replacement Vascular closure device 030212 general & internal medicine TRANSCATHETER Registries Aged 80 and over AORTIC-VALVE IMPLANTATION Endovascular Procedures General Medicine Aortic valvuloplasty REPLACEMENT Femoral Artery Treatment Outcome Female Cardiology and Cardiovascular Medicine Vascular Closure Devices medicine.medical_specialty Arteriotomy Hemorrhage Valvular and Structural Heart Diseases Transcatheter Aortic Valve Replacement 03 medical and health sciences Blood Vessel Prosthesis Implantation SDG 3 - Good Health and Well-being transcatheter valve implantation Catheterization Peripheral medicine Humans Radiology Nuclear Medicine and imaging Aged business.industry Hemostatic Techniques Aortic Valve Stenosis Surgery vascular closure Hemostasis 3121 General medicine internal medicine and other clinical medicine business MULTIDISCIPLINARY Aortic Aneurysm Abdominal |
Zdroj: | Catheterization and Cardiovascular Interventions Kroon, H G, Tonino, P A L, Savontaus, M, Amoroso, G, Laine, M, Christiansen, E H, Toggweiler, S, Ten Berg, J, Sathananthan, J, Daemen, J, de Jaegere, P P, Brueren, G B R G, Malmberg, M, Slagboom, T, Moriyama, N, Terkelsen, C J, Moccetti, F, Gheorghe, L, Webb, J, Wood, D & Van Mieghem, N M 2021, ' Dedicated plug based closure for large bore access-The MARVEL prospective registry ', Catheterization and Cardiovascular Interventions, vol. 97, no. 6, pp. 1270-1278 . https://doi.org/10.1002/ccd.29439 Catheterization and Cardiovascular Interventions, 97(6), 1270-1278. Wiley-Liss Inc. |
ISSN: | 1522-726X 1522-1946 |
DOI: | 10.1002/ccd.29439 |
Popis: | OBJECTIVES: To study safety and performance of the MANTA Vascular closure device (VCD) under real world conditions in 10 centers.BACKGROUND: The MANTA is a novel plug-based device for large bore arteriotomy closure.METHODS: We included all eligible patients who underwent transfemoral large bore percutaneous procedures. Exclusion criteria were per operator's discretion and included severe calcification or marked tortuosity of the access vessel, presence of marked obesity/cachexia or a systolic blood pressure above 180 mmHg. The primary performance endpoint was time to hemostasis. Primary and secondary safety endpoints were major and minor access site related vascular complications up to 30 days, respectively. Vascular complications were adjudicated by an independent clinical event committee according to VARC-2 criteria. We performed multivariable logistic regression to estimate the effect of baseline and procedural characteristics on any and major vascular complications.RESULTS: Between February 2018 and July 2019 500 patients were enrolled undergoing Transcatheter aortic valve replacement (TAVR, N = 496), Balloon aortic valvuloplasty (BAV, N = 2), Mechanical circulatory support (MCS, N = 1) or Endovascular aneurysm repair (EVAR, N = 1). Mean age was 80.8 ± 6.6 years with a median STS-score of 2.7 [IQR 2.0-4.3] %. MANTA access site complications were major in 20 (4%) and minor in 28 patients (5.6%). Median time to hemostasis was 50 [IQR 20-120] sec. Severe femoral artery calcification, scar presence in groin, longer procedure duration, female gender and history of hypertension were independent predictors for vascular complications.CONCLUSION: In this study, MANTA appeared to be a safe and effective device for large bore access closure under real-world conditions. |