Optimal frequency range for medical radar measurements of human heartbeats using body-contact radar

Autor: Sverre Brovoll, Yoann Paichard, Svein-Erik Hamran, Øyvind Aardal, Tor Berger, Tor Sverre Lande
Rok vydání: 2013
Předmět:
Zdroj: EMBC
ISSN: 2694-0604
Popis: In this paper the optimal frequency range for heartbeat measurements using body-contact radar is experimentally evaluated. A Body-contact radar senses electromagnetic waves that have penetrated the human body, but the range of frequencies that can be used are limited by the electric properties of the human tissue. The optimal frequency range is an important property needed for the design of body-contact radar systems for heartbeat measurements. In this study heartbeats are measured using three different antennas at discrete frequencies from 0.1 - 10 GHz, and the strength of the received heartbeat signal is calculated. To characterize the antennas, when in contact with the body, two port S-parameters(†) are measured for the antennas using a pork rib as a phantom for the human body. The results shows that frequencies up to 2.5 GHz can be used for heartbeat measurements with body-contact radar.
Databáze: OpenAIRE