Death of the Safety Guidewire
Autor: | Ralph V. Clayman, Rahul Dutta, Jaime Landman, Aashay Vyas |
---|---|
Rok vydání: | 2016 |
Předmět: |
Ureteral Perforation
medicine.medical_specialty Urethral meatus Percutaneous medicine.diagnostic_test business.industry Urology 030232 urology & nephrology Equipment Design History 20th Century Collection system History 21st Century Surgery 03 medical and health sciences 0302 clinical medicine 030220 oncology & carcinogenesis medicine Ureteroscopes Ureteroscopy Humans Safety business Urinary Catheterization |
Zdroj: | Journal of endourology. 30(9) |
ISSN: | 1557-900X |
Popis: | The safety guidewire has been an integral tool in endourologic percutaneous and ureteroscopic procedures for the last three decades, providing access to the collecting system or ureter in the event of loss of renal contact or ureteral perforation, respectively. However, recent advances in endourologic techniques and instrumentation have obviated the need of routine safety guidewire placement. Today the establishment of a "through-and-through" flank to urethral meatus guidewire eliminates the need of an additional guidewire during percutaneous procedures. Likewise, the availability of smaller flexible ureteroscopes, as well as modern ureteral access sheaths, has reduced the necessity of safety guidewire placement in ureteroscopic procedures. Herein, we trace the historical development of the safety guidewire concept, review recent advances in technologies that have obsoleted the safety guidewire, and evaluate recent data suggesting that continued use of a safety guidewire during ureteroscopic procedures may indeed be counterproductive. |