Imaging dose management using multi-resolution in CT-guided radiation therapy
Autor: | Ke Sheng, Thomas R. Mackie, Robert Jeraj, Rick Shaw, Bhudatt R. Paliwal |
---|---|
Rok vydání: | 2005 |
Předmět: |
Image quality
medicine.medical_treatment Radiation Dosage Models Biological Sensitivity and Specificity Tomotherapy Imaging Three-Dimensional Wavelet Region of interest medicine Humans Dosimetry Computer Simulation Radiology Nuclear Medicine and imaging Radiation Injuries Radiometry Radiological and Ultrasound Technology Radon transform business.industry Reproducibility of Results Radiotherapy Dosage Radiotherapy Computer-Assisted Radiographic Image Enhancement Radiation therapy Radiographic Image Interpretation Computer-Assisted Tomography Radiotherapy Conformal Tomography X-Ray Computed business Nuclear medicine Algorithms |
Zdroj: | Physics in Medicine and Biology. 50:1205-1219 |
ISSN: | 1361-6560 0031-9155 |
DOI: | 10.1088/0031-9155/50/6/011 |
Popis: | In image-guided radiation therapy, megavoltage computerized tomography (MVCT) delivers higher dose to the patient for lower image quality than diagnostic kilovoltage CT (kVCT). One way to reduce the mean imaging dose is to reduce the imaging volume, which is often sufficient for registration and dosimetry purposes. The filtered back projection using truncated data causes artefacts that degrade the image quality. Those artefacts can be effectively reduced by wavelet-based multi-resolution analysis (WMRA), in which the detail and approximate information are reconstructed separately to bypass the non-locality of filtered back projection. In this study, WMRA was used to reconstruct local images from both very low-dose kVCT scans from a bench-top tomotherapy unit and MVCT scans from helical tomotherapy. Results show that mean imaging dose can be significantly reduced by imaging a small region of interest. In simulation, the root-mean-square error brought by the truncation is smaller than 1-2% and depends on the level of dose reduction. On the other hand, the same mean dose that would have been delivered by a low-quality global CT can be conformed to a smaller volume to improve the visibility of low-contrast organs and fine structures using WMRA. Organs at risk can be avoided during repeated daily CT imaging when irregular-shaped reconstruction areas are used. WMRA does not involve computationally expensive iterations and is suitable for image-guided radiation therapy where imaging speed is essential. Compared with extrapolation methods, errors are further reduced to improve the detection of low contrast and fine structures. |
Databáze: | OpenAIRE |
Externí odkaz: |