Autor: |
Kaggie, Joshua D., Lanz, Titus, McLean, Mary A., Riemer, Frank, Schulte, Rolf F., Benjamin, Arnold J. V., Kessler, Dimitri A., Sun, Chang, Gilbert, Fiona J., Graves, Martin J., Gallagher, Ferdia A. |
Přispěvatelé: |
Kaggie, Joshua D. [0000-0001-6706-3442], McLean, Mary A. [0000-0002-3752-0179], Riemer, Frank [0000-0002-3805-5221], Benjamin, Arnold J. V. [0000-0003-2063-8258], Graves, Martin J. [0000-0003-4327-3052], Apollo - University of Cambridge Repository |
Rok vydání: |
2021 |
Předmět: |
|
Popis: |
Purpose: An unmet need in carbon‐13 (13C)‐MRI is a transmit system that provides uniform excitation across a large FOV and can accommodate patients of wide‐ranging body habitus. Due to the small difference between the resonant frequencies, sodium‐23 (23Na) coil developments can inform 13C coil design while being simpler to assess due to the higher naturally abundant 23Na signal. Here we present a removable 23Na birdcage, which also allows operation as a 13C abdominal coil. Methods: We demonstrate a quadrature‐driven 4‐rung 23Na birdcage coil of 50 cm in length for both 23Na and 13C abdominal imaging. The coil transmit efficiencies and B 1 + maps were compared to a linearly driven 13C Helmholtz‐based (clamshell) coil. SNR was investigated with 23Na and 13C data using an 8‐channel 13C receive array within the 23Na birdcage. Results: The 23Na birdcage longitudinal FOV was > 40 cm, whereas the 13C clamshell was < 32 cm. The transmit efficiency of the birdcage at the 23Na frequency was 0.65 µT/sqrt(W), similar to the clamshell for 13C. However, the coefficient of variation of 23Na‐ B 1 + was 16%, nearly half that with the 13C clamshell. The 8‐channel 13C receive array combined with the 23Na birdcage coil generated a greater than twofold increase in 23Na‐SNR from the central abdomen compared with the birdcage alone. Discussion: This 23Na birdcage coil has a larger FOV and improved B 1 + uniformity when compared to the widely used clamshell coil design while also providing similar transmit efficiency. The coil has the potential to be used for both 23Na and 13C imaging. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|