A green method for the quantification of plastics-derived endocrine disruptors in beverages by chemometrics-assisted liquid chromatography with simultaneous diode array and fluorescent detection
Autor: | Gabriela A. Ibañez, Rocio Pellegrino Vidal, Graciela M. Escandar |
---|---|
Rok vydání: | 2016 |
Předmět: |
Resolution (mass spectrometry)
Dibutyl phthalate Analytical chemistry BEVERAGES 02 engineering and technology Endocrine Disruptors Diethyl phthalate 01 natural sciences Fluorescence ENDOCRINE DISRUPTORS Analytical Chemistry Beverages Chemometrics chemistry.chemical_compound Detection limit Chromatography Chemistry Elution Otras Ciencias Químicas 010401 analytical chemistry Ciencias Químicas Phthalate 021001 nanoscience & nanotechnology LIQUID CHROMATOGRAPHY 0104 chemical sciences SECOND-ORDER CALIBRATION Gas chromatography–mass spectrometry 0210 nano-technology Plastics Food Analysis CIENCIAS NATURALES Y EXACTAS Chromatography Liquid |
Zdroj: | Talanta. 159:336-343 |
ISSN: | 0039-9140 |
DOI: | 10.1016/j.talanta.2016.06.049 |
Popis: | The aim of this study was to develop a novel analytical method for the determination of bisphenol A, nonylphenol, octylphenol, diethyl phthalate, dibutyl phthalate and diethylhexyl phthalate, compounds known for their endocrine-disruptor properties, based on liquid chromatography with simultaneous diode array and fluorescent detection. Following the principles of green analytical chemistry, solvent consumption and chromatographic run time were minimized. To deal with the resulting incomplete resolution in the chromatograms, a second-order calibration was proposed. Second-order data (elution time-absorbance wavelength and elution time-fluorescence emission wavelength matrices) were obtained and processed by multivariate curve resolution-alternating least-squares (MCR-ALS). Applying MCR-ALS allowed quantification of the analytes even in the presence of partially overlapped chromatographic and spectral bands among these compounds and the potential interferents. The obtained results from the analysis of beer, wine, soda, juice, water and distilled beverage samples were compared with gas chromatography-mass spectrometry (GC-MS). Limits of detection (LODs) in the range 0.04–0.38 ng mL−1 were estimated in real samples after a very simple solid-phase extraction. All the samples were found to contain at least three EDs, in concentrations as high as 334 ng mL−1. Fil: Pellegrino Vidal, Rocio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Rosario. Instituto de Química Rosario. Universidad Nacional de Rosario. Facultad de Ciencias Bioquímicas y Farmacéuticas. Instituto de Química Rosario; Argentina Fil: Ibañez, Gabriela Alejandra. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Rosario. Instituto de Química Rosario. Universidad Nacional de Rosario. Facultad de Ciencias Bioquímicas y Farmacéuticas. Instituto de Química Rosario; Argentina Fil: Escandar, Graciela Monica. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Rosario. Instituto de Química Rosario. Universidad Nacional de Rosario. Facultad de Ciencias Bioquímicas y Farmacéuticas. Instituto de Química Rosario; Argentina |
Databáze: | OpenAIRE |
Externí odkaz: |