HIGH FIDELITY FETAL RADIOELECTROCARDIOGRAPHY
Autor: | M. Maxim Asa, Arthur Bernstein, Edwin L. Rothfeld, I. Richard Zucker, Lewis E. Savel, Asa H. Crews |
---|---|
Rok vydání: | 1964 |
Předmět: |
medicine.medical_specialty
Fetus Labor Obstetric medicine.diagnostic_test business.industry Low-pass filter Active Labor Uterine contraction QRS complex Electrocardiography High fidelity Fetal Heart Diagnostic quality Equipment and Supplies Pregnancy Internal medicine medicine Cardiology Humans Telemetry Female medicine.symptom Cardiology and Cardiovascular Medicine business |
Zdroj: | The American journal of cardiology. 14 |
ISSN: | 0002-9149 |
Popis: | 1. 1. A new method for obtaining fetal electrocardiograms of superior quality has been described. 2. 2. Our technic employs a dual channel radioelectrocardiograph, a low pass filter, a miniaturized microamplifier electrode assembly, and a silver oxide conductive jelly developed in this laboratory. 3. 3. Fetal electrocardiograms of diagnostic quality were obtained in 47 of 50 consecutive patients studied during active labor. 4. 4. The tracings taken with this system have a stable baseline and no somatic tremor even during an active uterine contraction. The fetal QRS complexes are shown in excellent detail, and their voltage equals or exceeds that of the maternal. |
Databáze: | OpenAIRE |