Infection prevention and control in ultrasound - best practice recommendations from the European Society of Radiology Ultrasound Working Group

Autor: Paul S. Sidhu, Michel Claudon, Roland J. Koerner, Mirko D'Onofrio, Hilary Humphreys, Christiane M. Nyhsen, Afshin Gangi, Gerhard Mostbeck, Carlos Nicolau, Nicolas Grenier, Adrian P. Brady
Přispěvatelé: Groupe hospitalier Pellegrin, Laboratoire des sciences de l'ingénieur, de l'informatique et de l'imagerie (ICube), École Nationale du Génie de l'Eau et de l'Environnement de Strasbourg (ENGEES)-Université de Strasbourg (UNISTRA)-Institut National des Sciences Appliquées - Strasbourg (INSA Strasbourg), Institut National des Sciences Appliquées (INSA)-Institut National des Sciences Appliquées (INSA)-Institut National de Recherche en Informatique et en Automatique (Inria)-Les Hôpitaux Universitaires de Strasbourg (HUS)-Centre National de la Recherche Scientifique (CNRS)-Matériaux et Nanosciences Grand-Est (MNGE), Université de Strasbourg (UNISTRA)-Université de Haute-Alsace (UHA) Mulhouse - Colmar (Université de Haute-Alsace (UHA))-Institut National de la Santé et de la Recherche Médicale (INSERM)-Institut de Chimie du CNRS (INC)-Centre National de la Recherche Scientifique (CNRS)-Université de Strasbourg (UNISTRA)-Université de Haute-Alsace (UHA) Mulhouse - Colmar (Université de Haute-Alsace (UHA))-Institut National de la Santé et de la Recherche Médicale (INSERM)-Institut de Chimie du CNRS (INC)-Centre National de la Recherche Scientifique (CNRS)-Réseau nanophotonique et optique, Université de Strasbourg (UNISTRA)-Université de Haute-Alsace (UHA) Mulhouse - Colmar (Université de Haute-Alsace (UHA))-Centre National de la Recherche Scientifique (CNRS)-Université de Strasbourg (UNISTRA)-Centre National de la Recherche Scientifique (CNRS), Centre Hospitalier Régional Universitaire de Nancy (CHRU Nancy)
Rok vydání: 2017
Předmět:
Zdroj: Insights into Imaging, Vol 8, Iss 6, Pp 523-535 (2017)
Insights into Imaging
Insights into Imaging, 2017, 8 (6), pp.523-535. ⟨10.1007/s13244-017-0580-3⟩
ISSN: 1869-4101
DOI: 10.1007/s13244-017-0580-3
Popis: The objective of these recommendations is to highlight the importance of infection prevention and control in ultrasound (US), including diagnostic and interventional settings. Review of available publications and discussion within a multidisciplinary group consistent of radiologists and microbiologists, in consultation with European patient and industry representatives. Good basic hygiene standards are essential. All US equipment must be approved prior to first use, including hand held devices. Any equipment in direct patient contact must be cleaned and disinfected prior to first use and after every examination. Regular deep cleaning of the entire US machine and environment should be undertaken. Faulty transducers should not be used. As outlined in presented flowcharts, low level disinfection is sufficient for standard US on intact skin. For all other minor and major interventional procedures as well as all endo-cavity US, high level disinfection is mandatory. Dedicated transducer covers must be used when transducers are in contact with mucous membranes or body fluids and sterile gel should be used inside and outside covers. Good standards of basic hygiene and thorough decontamination of all US equipment as well as appropriate use of US gel and transducer covers are essential to keep patients safe. • Transducers must be cleaned/disinfected before first use and after every examination. • Low level disinfection is sufficient for standard US on intact skin. • High level disinfection is mandatory for endo-cavity US and all interventions. • Dedicated transducer covers must be used for endo-cavity US and all interventions. • Sterile gel should be used for all endo-cavity US and all interventions.
Databáze: OpenAIRE