Dose reduction potential in diagnostic single energy CT through patient‐specific prefilters and a wider range of tube voltages
Autor: | Joscha Maier, Marc Kachelrieß, Stefan Sawall, Jörg Steidel |
---|---|
Rok vydání: | 2021 |
Předmět: |
Adult
Materials science Drug Tapering Phantoms Imaging media_common.quotation_subject General Medicine Patient specific Radiation Dosage Imaging phantom Range (statistics) Humans Contrast (vision) Tube (fluid conveyance) Dose reduction Child Tomography X-Ray Computed Monte Carlo Method Energy (signal processing) media_common Voltage Biomedical engineering |
Zdroj: | Medical Physics. 49:93-106 |
ISSN: | 2473-4209 0094-2405 |
DOI: | 10.1002/mp.15355 |
Popis: | PURPOSE Various studies have demonstrated that additional prefilters and/or reduced tube voltages have the potential to significantly increase the contrast-to-noise ratios at unit dose (CNRDs) and thereby to significantly reduce patient dose in clinical CT. An exhaustive analysis, accounting for a wide range of filter thicknesses and a wide range of tube voltages extending beyond the 70 to 150 kV range of today's CT systems, including their specific choice depending on the patient size, is, however, missing. Therefore this work analyzes the dose reduction potential for patient-specific selectable prefilters combined with a wider range of tube voltages. We do so for soft tissue and iodine contrast in single energy CT. The findings may be helpful to guide further developments of x-ray tubes and automatic filter changers. METHODS CT acquisitions were simulated for different patient sizes (semi-anthropomorphic phantoms for child, adult and obese patients), tube voltages (35 to 150 kV), prefilter materials (tin and copper) and prefilter thicknesses (up to 5 mm). For each acquisition soft tissue and iodine CNRDs were determined. Dose was calculated using Monte Carlo simulations of a CTDI phantom. CNRD values of acquisitions with different parameters were used to evaluate dose reduction. RESULTS Dose reduction through patient-specific prefilters depends on patient size and available tube current among others. With an available tube current time product of 1000 mAs dose reductions of 17 % for the child, 32 % for the adult and 29 % for the obese phantom were achieved for soft tissue contrast. For iodine contrast dose reductions were 57 %, 49 % and 39 % for child, adult and obese phantoms, respectively. Here a tube voltage range extended to lower kV is important. CONCLUSIONS Substantial dose reduction can be achieved by utilizing patient-specific prefilters. Tube voltages lower than 70 kV are beneficial for dose reduction with iodine contrast, especially for small patients. The optimal implementation of patient-specific prefilters benefits from higher tube power. Tin prefilters should be available in 0.1 mm steps or lower, copper prefilter in 0.3 mm steps or lower. At least 10 different prefilter thicknesses should be used to cover the dose optima of all investigated patient sizes and contrast mechanisms. In many cases it would be advantageous to adapt the prefilter thicknes rather than the tube current to the patient size, i.e. to always use the maximum available tube current and to control the exposure by adjusting the thickness of the prefilter. This article is protected by copyright. All rights reserved. |
Databáze: | OpenAIRE |
Externí odkaz: |