Commutability assessment of potential reference materials using a multicenter split-patient-sample between-field-methods (twin-study) design: study within the framework of the Dutch project 'Calibration 2000'

Autor: Christa M. Cobbaert, Rob Jansen, H. Baadenhuijsen, Cas Weykamp, Aldy Kuypers, Herman Steigstra
Rok vydání: 2002
Předmět:
Zdroj: Clinical Chemistry, 48, 1520-5
Scopus-Elsevier
Clinical Chemistry, 48, 9, pp. 1520-5
Europe PubMed Central
ISSN: 0009-9147
Popis: Background: The Dutch project “Calibration 2000” aims at harmonization of laboratory results via calibration by development of commutable, matrix-based, secondary reference materials. An alternative approach to the NCCLS EP14 protocol for studying commutability of reference materials is presented, the “twin-study design”, which in essence is a multicenter, split-patient-sample, between-field-methods protocol. Methods: The study consisted of the simultaneous analysis of fresh patient sera and potential reference materials (PRMs) for HDL-cholesterol (HDL-C) by 86 laboratories forming 43 laboratory couples. Six subgroups of method combinations were formed. The patient sera were selected and interchanged by each laboratory couple. The PRMs consisted of three types: C37, prepared according to the NCCLS C37 protocol; Fro, frozen selectively pooled human serum; and Lyo, which was the same serum pool as Fro but lyophilized in the presence of sucrose. All PRMs were provided in three HDL-C concentrations. The regression line residuals for the PRMs were normalized by expressing them as multiples of the state-of-the-art within laboratory SD (SDSA). In addition, the extra contribution of each PRM to the total measurement uncertainty, CVNetto, was calculated. Results: Averaged over the three PRM concentrations, 1.6% of the C37 residuals were outside the 3 SDSA limit. For the Fro and Lyo PRMs, these values were 2.4% and 11.1%. CVNetto values for C37, Fro, and Lyo were 2.9%, 4.3%, and 5.3%, respectively. Conclusions: The present twin-study design, as a practical alternative to the NCCLS EP14 protocol, is a viable way of studying commutability characteristics of PRMs. The study suggests that the C37 PRMs are the best candidates for a future reference material.
Databáze: OpenAIRE