3 Tesla magnetic resonance imaging noise in standard head and neck sequence does not cause temporary threshold shift in high frequency
Autor: | Prepageran Narayanan, Mohammadreza Peyman, Elizabeth Yenn Lynn Lim, Ing Ping Tang, Raman Rajagopalan, Norlisah Ramli |
---|---|
Rok vydání: | 2014 |
Předmět: |
Adult
Male medicine.medical_specialty Adolescent Hearing loss Audiology Cohort Studies Young Adult Humans Medicine Aged 3 Tesla Magnetic Resonance Imaging Absolute threshold of hearing medicine.diagnostic_test business.industry Auditory Threshold Magnetic resonance imaging General Medicine Middle Aged Magnetic Resonance Imaging Noise Hearing Loss Noise-Induced Otorhinolaryngology Head and Neck Neoplasms Audiometry Pure-Tone Female Pure tone audiometry medicine.symptom business Nuclear medicine Auditory fatigue |
Zdroj: | European Archives of Oto-Rhino-Laryngology. 272:3109-3113 |
ISSN: | 1434-4726 0937-4477 |
DOI: | 10.1007/s00405-014-3232-y |
Popis: | High acoustic noise level is one of the unavoidable side effects of 3 T magnetic resonance imaging (MRI). A case of hearing loss after 3 T MRI has been reported in this institution and hence this study. The objective of this study was to determine whether temporary threshold shift (TTS) in high frequency hearing occurs in patients undergoing 3 T MRI scans of the head and neck. A total of 35 patients undergoing head and neck 3 T MRI for various clinical indications were tested with pure tone audiometry in different frequencies including high frequencies, before and after the MRI scan. Any threshold change from the recorded baseline of 10 dB was considered significant. All patients were fitted with foamed 3 M earplugs before the procedure following the safety guidelines for 3 T MRI. The mean time for MRI procedure was 1,672 s (range 1,040-2,810). The noise dose received by each patient amounted to an average of 3,906.29% (1,415-9,170%). The noise dose was derived from a normograph used by Occupational Noise Surveys. This was calculated using the nomograph of L eq, L EX, noise dose and time. There was no statistically significant difference between the hearing threshold before and after the MRI procedures for all the frequencies (paired t test, P > 0.05). For patients using 3 M foamed earplugs, noise level generated by 3 T MRI during routine clinical sequence did not cause any TTS in high frequency hearing. |
Databáze: | OpenAIRE |