A method to improve fluence resolution derived from two-dimensional detector array measurements for patient-specific IMRT verification using the information collected in dynalog files
Autor: | Maria Elisa Lavado Rodriguez, Juan Agustin Calama Santiago, Miguel Angel Infante Utrilla |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2015 |
Předmět: |
lcsh:Medical physics. Medical radiology. Nuclear medicine
Observational error Dynalog Computer science business.industry lcsh:R895-920 Resolution (electron density) Biophysics fluence Translation (geometry) intensity-modulated radiotherapy Fluence Imaging phantom 2D array Multileaf collimator Optics Histogram three-dimensional verification Radiology Nuclear Medicine and imaging Original Article Radiation treatment planning business |
Zdroj: | Journal of Medical Physics, Vol 40, Iss 1, Pp 5-12 (2015) Journal of Medical Physics / Association of Medical Physicists of India |
ISSN: | 1998-3913 0971-6203 |
Popis: | This paper proposes a method for improving the resolution of the fluence derived from detector array measurement using the information collected in dynalog files. From dynalog information, a file is generated with the actual multileaf collimator (MLC) positions and used as input to the treatment planning system (TPS) to obtain the dynalog-derived fluence and the theoretical response over the detector array. In contrast with the measured response, this theoretical response allows for correction of the dynalog-derived fluence and translation into the reconstructed fluence. This fluence is again introduced into the planning system to verify the treatment using clinical tools. Initially, more than 98% of the points passed the two-dimensional (2D) phantom gamma test (3% local dose - 3 mm) for all of the treatment verifications, but in some dose-volume histogram (DVH) comparisons, we note sensitive differences for the planning target volume (PTV) coverage and for the maximum doses in at-risk organs (up to 3.5%). In dose-distribution evaluations, we found differences of up to 5% in the PTV edges in certain cases due to detector array measurement errors. This work improves the resolution of the fluence derived from detector array measurements based on the treatment information, in contrast with the current commercial proposals based on planned data. |
Databáze: | OpenAIRE |
Externí odkaz: |