Autor: |
Levine Jh, Cesa M, Schecter So, Hoch Dh, Greenberg Sm |
Rok vydání: |
1997 |
Předmět: |
|
Zdroj: |
Pacing and Clinical Electrophysiology. 20:1721-1722 |
ISSN: |
1540-8159 0147-8389 |
DOI: |
10.1111/j.1540-8159.1997.tb03547.x |
Popis: |
Third generation implantable cardioverter defibrillators are capable of complex arrhythmia detection using sensing algorithms with automatic adjustable gain settings. We report a unique case where inappropriate sensing of atrial tachycardia in a patient with a His bundle ablation lead to satisfaction of ventricular fibrillation detection criteria. |