Inappropriate Discharges of an Implantable Cardioverter Defibrillator Secondary to Automatic Adjustable Gain of Atrial Tachycardia

Autor: Levine Jh, Cesa M, Schecter So, Hoch Dh, Greenberg Sm
Rok vydání: 1997
Předmět:
Zdroj: Pacing and Clinical Electrophysiology. 20:1721-1722
ISSN: 1540-8159
0147-8389
DOI: 10.1111/j.1540-8159.1997.tb03547.x
Popis: Third generation implantable cardioverter defibrillators are capable of complex arrhythmia detection using sensing algorithms with automatic adjustable gain settings. We report a unique case where inappropriate sensing of atrial tachycardia in a patient with a His bundle ablation lead to satisfaction of ventricular fibrillation detection criteria.