Efficacy of closed-loop stimulation with epicardial leads in an infant with congenital atrioventricular block
Autor: | Innocenzo Bianca, Salvatore Agati, Alfredo Di Pino |
---|---|
Rok vydání: | 2008 |
Předmět: |
Pacemaker
Artificial medicine.medical_specialty Atrial pacing business.industry Cardiac Pacing Artificial Infant Stimulation equipment and supplies medicine.disease Closed loop stimulation Electrocardiography Treatment Outcome Heart Rate Physiology (medical) Anesthesia Internal medicine medicine Cardiology Humans Atrioventricular Block Cardiology and Cardiovascular Medicine business Pericardium Atrioventricular block |
Zdroj: | Europace. 10:334-335 |
ISSN: | 1532-2092 1099-5129 |
DOI: | 10.1093/europace/eum299 |
Popis: | We present the case of an infant with a congenital atrioventricular block and sinus node dysfunction that was implanted with a pacemaker with physiological closed-loop stimulation (DDD-CLS) pacing mode with two epicardial leads. We have observed the response of DDD-CLS system with respect to conventional rate responsive accelerometer-based mode (DDDR): the atrial pacing percentage and the 24 h rate trend showed higher values during DDD-CLS compared with DDDR. In this patient, rate responsive feature seems preserved in DDD-CLS also using epicardial leads. |