High impedance alert with safety switching: An unreported hazard of hybrid pacing systems

Autor: Arlene Dermovsesian, Ankur A. Karnik, Matthew D. Gaskill, Kevin M. Monahan, Robert H. Helm
Rok vydání: 2019
Předmět:
Zdroj: J Cardiovasc Electrophysiol
ISSN: 1540-8167
1045-3873
DOI: 10.1111/jce.13941
Popis: Several Boston Scientific pacemaker models have a known issue with intermittent oversensing of the minute ventilation sensor when paired with non-Boston Scientific leads. Several of our patients with these hybrid systems have had transient out of range impedances and oversensing after safety switching which we suspected may be related. A retrospective analysis of 395 patients who had pacemakers implanted between 2015-2017 found that transient out of range impedances with safety switching was present in 9% of Boston Scientific pacemakers paired with Abbott or Medtronic leads compared with 0% in other device-lead combinations (P = 0.0089). We postulate that the root cause of the minute ventilation oversensing and transient high impedance issue is the same, a header-lead interaction from low-level incompatibility. Recognizing this issue is critical to prevent unnecessary lead revisions or extractions as it can be prevented with a simple reprogramming of lead pace/sense configuration.
Databáze: OpenAIRE