Load-Release Points of Two Novel Ureteral Stone-Trapping Devices
Autor: | Paulette Holley, Sameer K. Sharma, Michael K. Ouwenga, Kent T. Perry, Thomas M.T. Turk |
---|---|
Rok vydání: | 2005 |
Předmět: |
medicine.medical_specialty
Ureteral Calculi business.industry Urology Urinary stone Ureteral stone Equipment Design Surgical Instruments Models Biological In vitro model Surgery Bead (woodworking) Ureter medicine.anatomical_structure Materials Testing Ureteroscopy Humans Medicine business Biomedical engineering |
Zdroj: | Journal of Endourology. 19:894-897 |
ISSN: | 1557-900X 0892-7790 |
DOI: | 10.1089/end.2005.19.894 |
Popis: | Purpose: We compared the load-release points of the Stone Cone™ and the new Cook Urological NTrap™ devices using an in vitro model of the ureter. Materials and Methods: A strictured-ureter model was created using urethane tubing with a wire tie around it. A plastic bead was lodged above this narrowing. The Stone Cone, then the NTrap, was deployed in turn above this bead. The force required for each device to undeploy was measured over five trials. Results: Each device released the bead in every trial. The Stone Cone released at a mean of 0.190 lb, while the NTrap yielded at 0.861 lb. Conclusions: Both of the devices released the plastic bead reliably and at low loads, although the NTrap did require statistically significantly more force than the Stone Cone. However, the loads for both devices were |
Databáze: | OpenAIRE |
Externí odkaz: |