Male medicine.medical_specialty Biomedical Engineering Biophysics Bioengineering 030204 cardiovascular system & hematology Biomaterials 03 medical and health sciences 0302 clinical medicine Internal medicine Medicine Humans Postoperative Period Thrombus Heart-Assist Devices Pump thrombosis Heart Failure business.industry Continuous flow Anticoagulants Thrombosis General Medicine medicine.disease 030228 respiratory system Heart failure Acute Disease cardiovascular system Cardiology business |
Zdroj: | ASAIO Journal, 65(7), E72-E74. Lippincott Williams & Wilkins |
ISSN: | 1538-943X 1058-2916 |
Popis: | There are no reports of acute pump thrombosis in the latest, continuous flow left ventricular assist devices type HeartMate 3, other than thrombus ingestion. We present a case of early thrombosis of the pump and outflow graft, necessitating acute pump and outflow graft replacement. A combination of lowflow episodes and subtherapeutic levels of anticoagulation was the most likely cause. ASAIO Journal 2019; 65:e72–e74. |