If hip implant retrievals could speak, what would they tell us?
Autor: | J. M. Cuckler |
---|---|
Rok vydání: | 2012 |
Předmět: |
Orthodontics
Third body medicine.medical_specialty Bearing (mechanical) business.industry Arthroplasty Replacement Hip Equipment Failure Analysis Fatigue testing Fretting Bone cement law.invention Surgery Prosthesis Failure Hip implant law Medicine Humans Orthopedics and Sports Medicine Implant Hip Prosthesis business Device Removal |
Zdroj: | The Journal of bone and joint surgery. British volume. 94(11 Suppl) |
ISSN: | 2044-5377 |
Popis: | Hip implant retrieval analysis is the most important source of insight into the performance of new materials and designs of hip arthroplasties. Even the most rigorous in vitro testing will not accurately simulate the behavior of implant materials and new designs of prosthetic arthroplasties. Retrieval analysis has revealed such factors as the effects of gamma-in-air sterilisation of polyethylene, fatigue failure mechanisms of polymethylmethacrylate bone cement, fretting corrosion of Morse taper junctions, third body wear effects of both hard-on-hard and hard-on-soft bearing couples, and the effects of impingement of components on the full spectrum of bearing surfaces, none of which was predicted by pre-implantation in vitro testing of these materials and combinations. The temporal sequence of the retrieval process is approximately six years from first implantation through retrieval analysis, laboratory investigation, and publication of results, and thus, in addition to rigorous clinical evaluation, represents the true development and insight cycle for new designs and materials. |