The Highly Miniaturised Radiation Monitor

Autor: S. Gunes-Lasnet, Alessandra Menicucci, Renato Turchetta, D. Griffin, E. Daly, S. Woodward, A. Marshall, Henrique Araujo, Nicola Guerrini, O. Poyntz-Wright, E. F. Mitchell, T. Morse
Rok vydání: 2014
Předmět:
DOI: 10.48550/arxiv.1401.3533
Popis: We present the design and preliminary calibration results of a novel highly miniaturised particle radiation monitor (HMRM) for spacecraft use. The HMRM device comprises a telescopic configuration of active pixel sensors enclosed in a titanium shield, with an estimated total mass of 52 g and volume of 15 cm$^3$. The monitor is intended to provide real-time dosimetry and identification of energetic charged particles in fluxes of up to 10$^8$ cm$^{-2}$ s$^{-1}$ (omnidirectional). Achieving this capability with such a small instrument could open new prospects for radiation detection in space.
Comment: 17 pages, 15 figures
Databáze: OpenAIRE