Dinamap fails to detect hypotension in very low birthweight infants

Autor: G K Diprose, M I Levene, David H. Evans, L N Archer
Rok vydání: 1986
Předmět:
Zdroj: Archives of Disease in Childhood. 61:771-773
ISSN: 1468-2044
0003-9888
DOI: 10.1136/adc.61.8.771
Popis: The accuracy of blood pressure measurements obtained from very low birthweight (less than 1500 g) neonates using the Dinamap oscillometric monitor was investigated. Comparisons using umbilical artery measurements showed that the monitor is less reliable in the lower pressure range and specifically that it tends to overestimate pressure in hypotensive infants.