A phantom for verification of dwell position and time of a high dose rate brachytherapy source
Autor: | Tomas Kron, Rick Franich, M. Madebo, Joseph Pillainayagam |
---|---|
Rok vydání: | 2012 |
Předmět: |
Materials science
medicine.medical_treatment Brachytherapy Biomedical Engineering Biophysics General Physics and Astronomy STRIPS Curvature Sensitivity and Specificity Imaging phantom law.invention law medicine Radiology Nuclear Medicine and imaging Computed radiography Radiometry Simulation Reproducibility Phantoms Imaging Detector Reproducibility of Results Radiotherapy Dosage Equipment Design equipment and supplies Equipment Failure Analysis Dwell time Tomography X-Ray Computed human activities Biomedical engineering |
Zdroj: | Australasian Physical & Engineering Sciences in Medicine. 35:335-339 |
ISSN: | 1879-5447 0158-9938 |
Popis: | Accuracy of dwell position and reproducibility of dwell time are critical in high dose rate (HDR) brachytherapy. A phantom was designed to verify dwell position and dwell time reproducibility for an Ir-192 HDR stepping source using Computed Radiography (CR). The central part of the phantom, incorporating thin alternating strips of lead and acrylic, was used to measure dwell positions. The outer part of the phantom features recesses containing different absorber materials (lead, aluminium, acrylic and polystyrene foam), and was used for determining reproducibility of dwell times. Dwell position errors of |
Databáze: | OpenAIRE |