A phantom for verification of dwell position and time of a high dose rate brachytherapy source

Autor: Tomas Kron, Rick Franich, M. Madebo, Joseph Pillainayagam
Rok vydání: 2012
Předmět:
Zdroj: Australasian Physical & Engineering Sciences in Medicine. 35:335-339
ISSN: 1879-5447
0158-9938
Popis: Accuracy of dwell position and reproducibility of dwell time are critical in high dose rate (HDR) brachytherapy. A phantom was designed to verify dwell position and dwell time reproducibility for an Ir-192 HDR stepping source using Computed Radiography (CR). The central part of the phantom, incorporating thin alternating strips of lead and acrylic, was used to measure dwell positions. The outer part of the phantom features recesses containing different absorber materials (lead, aluminium, acrylic and polystyrene foam), and was used for determining reproducibility of dwell times. Dwell position errors of
Databáze: OpenAIRE