An experimental set-up to test heat-moisture exchangers

Autor: P. E. M. Huygen, K. Jabaaij, Can Ince, I. Gultuna, B. Saygin, Hajo A. Bruining, Jan C. Pompe, Necmettin Ünal, Wim P.J. Holland
Přispěvatelé: Other departments
Jazyk: angličtina
Rok vydání: 1995
Předmět:
Zdroj: Intensive care medicine, 21(2), 142-148. Springer Verlag
ISSN: 0342-4642
DOI: 10.1007/bf01726537
Popis: The purpose of this study was to build an experimental set-up to assess continuously the humidification, heating and resistance properties of heat-moisture exchangers (HMEs) under clinical conditions. The experimental set-up consists of a patient model, measurement systems and a ventilator. Surgical ICU, University Hospital of Rotterdam. A clinically used HME. The air flow, pressure in the ventilation circuit, pressure difference over the HME, and partial water vapour pressure and temperature at each side of the HME were measured. The resistance, absolute humidity, humidification efficiency and temperature difference at the patient side of the HME were calculated. Measurements were performed during 24 h. The temperature output, humidity output and lung mechanics of the patient model were similar to values found in mechanically ventilated patients. The measurement system was in agreement with the ISO draft standard and was capable of measuring dynamic variation of water and heat exchange over the range of a clinically used ventilator setting. The experimental set-up described is reliable for evaluating HMEs and can also be used for future clinical evaluation of HMEs. The main advantages of this set-up over those described previously are: (i) measurements of dynamic variations of water and heat exchange; (ii) on-line measurements of expiratory, as well as inspiratory resistance
Databáze: OpenAIRE