Inappropriate ventricular fibrillation annotation and defibrillator discharge despite the same tachycardia cycle length in ventricular tachycardia‐1 zone in an Abbott defibrillator – What is the mechanism?
Autor: | Debabrata Bera, Siddhartha Mani, Ranabrata Sen, Suchit Majumder, Antareep Halder, Adrita Ray |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | Pacing and Clinical Electrophysiology. 46:169-171 |
ISSN: | 1540-8159 0147-8389 |
DOI: | 10.1111/pace.14654 |
Popis: | Newer devices incorporate advanced algorithms to have more accurate delivery of implantable cardioverter-defibrillator therapy. Newer models of Abbott Implantable cardioverter-defibrillators like Gallant and Entrant have a new ventricular fibrillation therapy assurance algorithm. This algorithm has a proven role in avoiding under-detection of ventricular fibrillation where some of the smaller electrograms are under-sensed. We report the first case of ventricular fibrillation therapy assurance algorithm malfunction in a single chamber Entrant Implantable cardioverter-defibrillator, where the patient received inappropriate therapy. The ventricular fibrillation therapy assurance algorithm eventually had to be turned off. This article is protected by copyright. All rights reserved. |