Development of a software tool using deterministic logic for the optimization of cochlear implant processor programming
Autor: | Bart Vaerenberg, Geert De Ceulaer, Carina De Beukelaer, K. Daemers, Paul J. Govaerts, Karen Schauwers |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2010 |
Předmět: |
Logic
Process (engineering) medicine.medical_treatment Expert Systems Prosthesis Design computer.software_genre Machine learning Outcome (game theory) Young Adult Intelligent agent Software Audiometry Cochlear implant Reflex Evoked Potentials Auditory Brain Stem Test suite Humans Medicine Psychoacoustics business.industry Stapedius Sensory Systems Expert system Cochlear Implants Otorhinolaryngology Evoked Potentials Auditory Speech Perception Female Neurology (clinical) Artificial intelligence Human medicine Electronics business computer Algorithms |
Zdroj: | Otology and neurotology |
ISSN: | 1531-7129 |
Popis: | Objective: An intelligent agent, Fitting to Outcomes eXpert, was developed to optimize and automate Cochlear implant (CI) programming. The current article describes the rationale, development, and features of this tool. Background: Cochlear implant fitting is a time-consuming procedure to define the value of a subset of the available electric parameters based primarily on behavioral responses. It is comfort-driven with high intraindividual and interindividual variability both with respect to the patient and to the clinician. Its validity in terms of process control can be questioned. Good clinical practice would require an outcome-driven approach. An intelligent agent may help solve the complexity of addressing more electric parameters based on a range of outcome measures. Methods: A software application was developed that consists of deterministic rules that analyze the map settings in the processor together with psychoacoustic test results (audiogram, A§E phoneme discrimination, A§E loudness scaling, speech audiogram) obtained with that map. The rules were based on the daily clinical practice and the expertise of the CI programmers. The data transfer to and from this agent is either manual or through seamless digital communication with the CI fitting database and the psychoacoustic test suite. It recommends and executes modifications to the map settings to improve the outcome. Results: Fitting to Outcomes eXpert is an operational intelligent agent, the principles of which are described. Its development and modes of operation are outlined, and a case example is given. Fitting to Outcomes eXpert is in use for more than a year now and seems to be capable to improve the measured outcome. Conclusion: It is argued that this novel tool allows a systematic approach focusing on outcome, reducing the fitting time, and improving the quality of fitting. It introduces principles of artificial intelligence in the process of CI fitting. Key Words: Artificial intelligenceVCochlear implantVDeterministic logicVFittingV Intelligent agentVOptimizationVOutcomeVProgrammingV Psychoacoustics. Otol Neurotol 31:908Y918, 2010. |
Databáze: | OpenAIRE |
Externí odkaz: |