Autor: |
L. S. August, C. C. Rogers, R. B. Theus, Frank H. Attix, P. Shapiro |
Rok vydání: |
1976 |
Předmět: |
|
Zdroj: |
Medical Physics. 3:87-90 |
ISSN: |
0094-2405 |
DOI: |
10.1118/1.594208 |
Popis: |
The displacement correction factor to be used for analysis of fast- neutron dosimetric measurements using air-filled EG and G tissue-equivalent ion chambers in a tissue-equivalent phantom has been investigated using the MANTA neutron radiotherapy beam generated by 35-MeV deuterons on a thick Be target. The displacement correction factor inferred from these measurements is 0.970 for the EG and G IC-17 (1.0-cm$sup 3$) ion chamber, and is 0.989 for the EG and G IC- 18 (0.1-cm$sup 3$) ion chamber. This multiplicative displacement correction factor has no significant dependence on depth in the phantom or on neutron beam size. (AIP) |
Databáze: |
OpenAIRE |
Externí odkaz: |
|