Effect of intrinsic background radiation on the operation of a Fourier spectroradiometer
Autor: | Igor V. Kochikov, S. K. Dvoruk, I. N. Efimov, Andrey N. Morozov, P. E. Shlygin, M. V. Lel’kov, S. I. Svetlichny, V. A. Pozdnyakov, S. E. Tabalin, V. N. Kornienko |
---|---|
Předmět: |
Physics
Spectrometer business.industry Applied Mathematics Astrophysics::Instrumentation and Methods for Astrophysics General Engineering Michelson interferometer Atomic and Molecular Physics and Optics law.invention Computational Mathematics Interferometry symbols.namesake Spectroradiometer Fourier transform Optics law Radiance symbols business Background radiation |
Zdroj: | Scopus-Elsevier |
Popis: | This paper discusses how the intrinsic background radiation of a Fourier spectroradiometer affects the results of the processing of two-sided interferograms. The operation of a Michelson interferometer is considered theoretically, and it is shown that, when objects with radiance temperatures close to the intrinsic temperature of the spectrometer are observed, inverted spectral sections and bands may appear in the final spectrum. The paper proposes an algorithm for correcting the initial interferograms and an experimental technique that makes it possible to eliminate the inversion of the spectrum and to compensate the intrinsic background radiation of the interferometer. |
Databáze: | OpenAIRE |
Externí odkaz: |