ATHENA X-IFU Demonstration Model: First Joint Operation of the Main TES Array and its Cryogenic AntiCoincidence Detector (CryoAC)

Autor: M. D’Andrea, K. Ravensberg, A. Argan, D. Brienza, S. Lotti, C. Macculi, G. Minervini, L. Piro, G. Torrioli, F. Chiarello, L. Ferrari Barusso, M. Biasotti, G. Gallucci, F. Gatti, M. Rigano, H. Akamatsu, J. Dercksen, L. Gottardi, F. de Groote, R. den Hartog, J.-W. den Herder, R. Hoogeveen, B. Jackson, A. McCalden, S. Rosman, E. Taralli, D. Vaccaro, M. de Wit, J. Chervenak, S. Smith, N. Wakeham
Rok vydání: 2022
Předmět:
Zdroj: Journal of Low Temperature Physics. 209:433-440
ISSN: 1573-7357
0022-2291
Popis: The X-IFU is the cryogenic spectrometer onboard the future ATHENA X-ray observatory. It is based on a large array of TES microcalorimeters, which works in combination with a Cryogenic AntiCoincidence detector (CryoAC). This is necessary to reduce the particle background level thus enabling part of the mission science goals. Here we present the first joint test of X-IFU TES array and CryoAC Demonstration Models, performed in a FDM setup. We show that it is possible to operate properly both detectors, and we provide a preliminary demonstration of the anti-coincidence capability of the system achieved by the simultaneous detection of cosmic muons.
Accepted for publication in the Journal of Low Temperature Physics for LTD-19 special issue
Databáze: OpenAIRE