Maximum respiratory pressure measuring system : calibration and evaluation of uncertainty

Autor: Verônica Franco Parreira, Carlos Julio Tierra-Criollo, JL Jose Ferreira, Marconi Oliveira, Nadja C. Pereira, Flavio H. Vasconcelos
Přispěvatelé: Signal Processing Systems
Jazyk: angličtina
Rok vydání: 2010
Předmět:
Zdroj: Revista Controle & Automação, 21(6), 588-597. Sociedade Brasileira de Automática (SBA)
Sba: Controle & Automação Sociedade Brasileira de Automatica v.21 n.6 2010
Sba: Controle & Automação Sociedade Brasileira de Automatica
Sociedade Brasileira de Automática (SBA)
instacron:SBA
ISSN: 0103-1759
DOI: 10.1590/s0103-17592010000600004
Popis: The objective of this paper is to present a methodology for the evaluation of uncertainties in the measurements results obtained during the calibration of a digital manovacuometer prototype (DM) with a load cell sensor pressure device incorporated. Calibration curves were obtained for both pressure sensors of the DM using linear regression by weighted least squares method (WLS). Two models were built to evaluate uncertainty. One takes into account the information listed in the sensor datasheet, resulting in the maximum permissible measurement error of the manovacuometer, and the other on the WLS implemented during calibration. Considering a range of ten calibration points, it was found that calibration procedure designed using WLS modeling indicates that the range of measurement uncertainty extends from 0.2 up to 0.5 kPa. This is inside the manufacter range that extends from 1.5 up 3.5 kPa, showing adequacy for use