Upgrade of the TAMU MDM-focal plane detector with MicroMegas technology
Autor: | R. Chyzh, A. Spiridon, Gheorghe Pascovici, M. Dag, E. C. Pollacco, Antti Saastamoinen, R. E. Tribble, Brian Roeder, R. De Oliveira, Livius Trache |
---|---|
Rok vydání: | 2016 |
Předmět: |
Radioactive ion beams
Physics Nuclear and High Energy Physics Spectrometer Physics::Instrumentation and Detectors 010308 nuclear & particles physics business.industry MicroMegas detector Focal plane detector 01 natural sciences Particle identification Nuclear physics Upgrade Optics Ionization 0103 physical sciences Ionization chamber 010306 general physics business Instrumentation |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 376:364-368 |
ISSN: | 0168-583X |
DOI: | 10.1016/j.nimb.2016.02.011 |
Popis: | A gridded ionization chamber used as a focal plane detector at the back of the TAMU-MDM spectrometer was modified to use MicroMegas technology for the purpose of improving energy resolution and particle identification. The upgraded system was tested in experimental conditions with several heavy-ion beams at 12 MeV/u and found to achieve resolutions between 3.2% and 4.8%. This is a significant improvement over the previous performance of 10–15% obtained using the existing, conventional ionization chambers. |
Databáze: | OpenAIRE |
Externí odkaz: |