Performance of a high sensitivity PET scanner based on LSO panel detectors
Autor: | Bernard Bendriem, Merence Sibomana, Michael E. Casey, Christian Michel, Lars Eriksson, C. Nahmias, Vladimir Y. Panin, David W. Townsend, V. Rappoport, W.F. Jones, Judson Jones, Maurizio Conti, Bjoern Jakoby |
---|---|
Rok vydání: | 2006 |
Předmět: |
Physics
Nuclear and High Energy Physics Noise measurement medicine.diagnostic_test business.industry Computer science Detector Iterative reconstruction Noise Equivalent Count Rate Line source Optics Nuclear Energy and Engineering Ordered subset expectation maximization Positron emission tomography Pet scanner medicine Tomography Sensitivity (control systems) Electrical and Electronic Engineering business Sensitivity (electronics) Image resolution |
Zdroj: | IEEE Nuclear Science Symposium Conference Record, 2005. |
DOI: | 10.1109/nssmic.2005.1596848 |
Popis: | We present the results of performance measurements for a new rotating panel detector research tomograph. The tomograph consists of five large-area lutetium oxyorthosilicate (LSO) panel detectors. A long axial field of view (53 cm), high sensitivity, and high count rate capability (due to recently developed fast electronics) are the most important features of this design. A peak noise equivalent count rate (NEC) of about 184 kcps has been measured, together with 2% absolute sensitivity for a 70 cm /sup 18/F line source and a spatial resolution less than 5 mm. Whole body patient images have been obtained, in two bed positions, 10 minutes per bed. An Ordinary Poisson Ordered Subset Expectation Maximization 3D algorithm (OP-OSEM3D) has been implemented on a dedicated computer cluster for reconstruction. |