Comparison of cervical motion restriction and interface pressure between two cervical collars

Autor: John D. McCamley, Marc C. Jacofsky, Kasra Araghi
Rok vydání: 2021
Předmět:
Zdroj: Clinical Biomechanics. 89:105482
ISSN: 0268-0033
DOI: 10.1016/j.clinbiomech.2021.105482
Popis: Background Cervical collar treatment is widespread amongst a broad range of ages and conditions. However, these devices are associated with a known correlation between collar usage and adverse effects such as pressure ulcers, infections, exacerbated spinal injury, and higher morbidity. The objective of this study was to determine if the ability of a newly developed DJO Procare XTEND 174 collar to restrict cervical spine motion was similar to that of the previously studied Ossur Miami J collar and to determine if this was done while producing similar tissue interface pressures. Methods Three-dimensional kinematic data were obtained for twelve healthy volunteer participants (6 female, 6 male) using a 10-camera infrared motion capture system. Cervical range of motion in each plane was calculated using angles between head and thorax rigid-body axes. Tissue interface pressure was measured between the head and the collar with three flexible pressure sensor pads over the anterior mandibles and occiput. The distribution of interface pressures was obtained in both the seated and supine positions. Findings Both collars significantly restricted range of motion in all movement directions (p