Falsely low values of oxygen saturation measured by pulse oximetry in a boy treated with Chinese herb tea
Autor: | Josef Briegel, Patricia Bernasconi, Aurelia Peraud, Tanija K. Hüttl, Agnes S. Meidert, Georg Hennig, Alexander Lang |
---|---|
Rok vydání: | 2016 |
Předmět: |
Male
Health Informatics Normal oxygen saturation Critical Care and Intensive Care Medicine Low oxygen saturation 03 medical and health sciences 0302 clinical medicine 030202 anesthesiology medicine Humans Arterial blood gas analysis Anesthesia Oximetry Child Blood gas analysis medicine.diagnostic_test Inpatient stay business.industry Reproducibility of Results 030208 emergency & critical care medicine Oxygenation medicine.disease Oxygen Pulse oximetry Anesthesiology and Pain Medicine Plant Preparations Blood Gas Analysis business Teas Herbal Drugs Chinese Herbal Glioblastoma |
Zdroj: | Journal of Clinical Monitoring and Computing. 31:481-484 |
ISSN: | 1573-2614 1387-1307 |
Popis: | An 8-year-old boy suffering from progressive glioblastoma was scheduled for neurosurgery. Prior to induction of anaesthesia pulse oximetry measured 64 % saturation of oxygen (SpO2). Arterial blood gas analysis revealed normal oxygen saturation and normal oxygen partial pressure. After having ruled out technical problems of pulse oximetry the neurosurgical procedure was halted. Meticulous examination of the child's history and medication did not explain a possible interaction of drugs with pulse oximetry. A Chinese herb tea had been given to the child, but was then stopped on the day of admission. The surgical procedure took place the next day without any complications. During the subsequent inpatient stay, repeated blood gas analyses showed normal oxygenation, but pulse oximetry measured initially SpO2 values of 64 %, gradually increasing over 7 days up to 91 % by the time of discharge from hospital. Blood samples were taken and analysed. Absorption spectroscopy from the patient's blood showed an uncommon absorption maximum at 684 nm besides the normal maxima. The normalisation of SpO2 values after stopping Chinese herb tea administration leads to the conclusion that one of its ingredients caused the distorted pulse oximetry measurement. |
Databáze: | OpenAIRE |
Externí odkaz: |