Aortic root morphology can cause early bioprosthetic valve failure of externally mounted pericardial bioprosthesis

Autor: Tomoki Ushijima, Naoto Shingu, Akira Shiose, Yoshihisa Tanoue, Yusuke Nakata, Sho Matsuyama, Satoshi Fujita, Hiromichi Sonoda, Yasuhisa Oishi, Satoshi Kimura
Rok vydání: 2020
Předmět:
Zdroj: General Thoracic and Cardiovascular Surgery. 69:107-109
ISSN: 1863-6713
1863-6705
Popis: A 73-year-old woman with a narrow aortic root had undergone aortic valve replacement with a 19-mm Mitroflow valve. Aortic annular enlargement with Manouguian's technique was performed, and the bioprosthesis had been implanted on the tilt in a supra-annular position. Four years after the implantation, echocardiography showed a significant de novo aortic regurgitation, which had not been detected 1 year earlier. In the reoperation, the left coronary leaflet of the bioprosthesis had attached and fused to the wall of the sinus of Valsalva, causing deformation of the valve leaflet and a commissural gap between the left and right coronary leaflets, which appeared to have caused the de novo aortic regurgitation. We assessed the cause of early valve deterioration by focusing on the morphology of the aortic root. Preoperative understanding of the aortic root morphology would help to avoid early valve dysfunction for aortic valve replacement with an externally mounted bioprosthesis.