Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice
Autor: | Dimitrios Papakonstadinou, Eleni Patsea, Katerina Chatzimichali, Sonia Chalkidou, Georgios Amariotakis, George Theodossiadis, Apostolos Amariotakis, Gerasimos Georgopoulos, Sotiris Skouriotis, Ioannis Halkiadakis |
---|---|
Rok vydání: | 2009 |
Předmět: |
Male
Intraocular pressure medicine.medical_specialty genetic structures Glaucoma Ocular hypertension Mean difference Goldmann applanation tonometry Tonometry Ocular Ophthalmology medicine Humans Intraocular Pressure Aged Aged 80 and over business.industry Reproducibility of Results General Medicine Middle Aged medicine.disease eye diseases body regions Female Ocular Hypertension sense organs business |
Zdroj: | Acta Ophthalmologica. 87:323-328 |
ISSN: | 1755-3768 1755-375X |
DOI: | 10.1111/j.1755-3768.2008.01239.x |
Popis: | Purpose: To compare intraocular pressure (IOP) readings taken using dynamic contour tonometry (DCT) with IOP readings taken with Goldmann applanation tonometry (GAT) in eyes with glaucoma or ocular hypertension. Methods: The present study included 100 eyes in 100 patients with glaucoma or ocular hypertension. After pachymetry DCT and GAT were performed. Intraocular pressures as measured with DCT and GAT were compared with one another and with central corneal thickness (CCT). Results: Mean DCT IOP measurements (20.1 ± 4.3 mmHg) were significantly (p |