Two miniaturised TEPCS in a single detector for BNCT microdosimetry
Autor: | V. Conte, L. De Nardo, D. Moro, Massimo Masi, G. Tornielli, P. Colautti, G. Gualdrini |
---|---|
Rok vydání: | 2006 |
Předmět: |
Photon
Transducers Boron Neutron Capture Therapy Radiation Boron neutron capture therapy Microdosimetry with tissue-equivalent proportional counters TEPC calibration in secondary standard photon fields Sensitivity and Specificity Photon field Nuclear magnetic resonance Optics Dosimetry Radiology Nuclear Medicine and imaging Radiometry Neutrons Physics Miniaturization Radiological and Ultrasound Technology business.industry Radiotherapy Planning Computer-Assisted Detector Public Health Environmental and Occupational Health Reproducibility of Results Equipment Design General Medicine Equipment Failure Analysis Systems Integration Neutron capture Volume (thermodynamics) Computer-Aided Design business Counting rate |
Zdroj: | Radiation Protection Dosimetry. 122:396-400 |
ISSN: | 1742-3406 0144-8420 |
DOI: | 10.1093/rpd/ncl484 |
Popis: | Microdosimetry with tissue-equivalent proportional counters (TEPC) has proven to be an ideal dosimetry technique for mixed radiation fields as those ones used in boron neutron capture therapy (BNCT). A new counter, composed of two twin cylindrical mini TEPCs inserted in a slim titanium sleeve of 2.7 mm external diameter, has been constructed. The detector has been designed to perform dosimetry and microdosimetry in intense radiation fields. The two mini TEPCs work in gas flow mode. They have right cylinder sensitive volumes of 0.9 mm. In spite of gas line tiny sizes, the gas pressure inside the two counters is well established with |
Databáze: | OpenAIRE |