Popis: |
Background: Actually, resonance frequency analysis (RFA) is the most extended method for measuring implant stability. The implant stability quotient (ISQ) is the measure obtained by the different RFA devices, however, inter- and intra- rater reliability and validity of some devices remains unknown. Methods: Thirty implants were placed in 3 different pig mandibles. ISQ was measured axial and parallel with Osstell® Beacon, Penguin® and MegaISQ® by 2 different operators and one operator performed a test-retest. Intraclass correlation coefficient was calculated to assess the intra- and inter-rater reliability. Pearson correlation coefficient was used to assess the validity. Results: The higher inter- and intra- rater reliability was obtained by Penguin® when measuring axial. The highest ISQ values were obtained using Penguin® in an axial measurement; the lowest, using the MegaISQ® in an axial measurement. The highest correlation values with the other devices were obtained by MegaISQ® measuring axially. Conclusion: Penguin® had a good reliability for measuring ISQ both inter- and intra- rater. Osstell® had good validity for measuring ISQ both axial and parallel and MegaISQ® had the best validity for measuring ISQ axial. |