The importance of low level QC for high sensitivity troponin assays
Autor: | Carmen Oakman, Tony Badrick, Peter E. Hickman, Gus Koerbin, Julia M. Potter |
---|---|
Rok vydání: | 2018 |
Předmět: |
Adult
Male Quality Control Detection limit Chromatography biology Chemistry Troponin I Clinical Biochemistry General Medicine 030204 cardiovascular system & hematology Sensitivity and Specificity Troponin 03 medical and health sciences 0302 clinical medicine 030220 oncology & carcinogenesis High sensitivity troponin biology.protein Humans Female Control material Blood Chemical Analysis Volume concentration |
Zdroj: | Clinical Biochemistry. 58:60-63 |
ISSN: | 0009-9120 |
DOI: | 10.1016/j.clinbiochem.2018.05.007 |
Popis: | Background With the advent of the new high-sensitivity troponin assays, it is becoming critical to measure troponin accurately to low concentrations. To ensure assay performance is acceptable, appropriate QC must be run. Methods In addition to the routine use of commercial QC material, we prepared pools of human QC material with low troponin concentrations close to the limit of quantitation, and ran these regularly on our laboratory analysers. Results Over 3 years we found no drift or shift in our hs-cTnI assay. We found that only the very low concentration human QC material gave warning of precision problems with the hs-cTnI assay. At the time of the documented poor assay precision, the higher concentration QC material indicated satisfactory performance. Conclusions Choice of QC material with an appropriate concentration is important for any assay. For hs-cTn assays, it is of particular importance to use control material with a concentration near to the limit of quantitation. |