Autor: |
Farhad Abtahi, Gudmundur Johannsson, L. Ellegard, Shirin Abtahi, Fernando Seoane, Ingvar Bosaeus, Leigh C. Ward |
Rok vydání: |
2015 |
Předmět: |
|
Zdroj: |
EMBC |
DOI: |
10.1109/embc.2015.7319198 |
Popis: |
Electrical bioimpedance has been used for several decades to assess body fluid distribution and body composition by using single frequency and bioimpedance spectroscopic (BIS) techniques. It remains uncertain whether BIS methods have better performance compare to single frequency regression equations. In this work the performance of two BIS methods and four different 50 kHz single frequency prediction equations was studied in a data set of wrist-to-ankle tetrapolar BIS measurements (5-1000 kHz) together with reference values of total body water obtained by tritium dilution in 92 patients. Data were compared using regression techniques and Bland-Altman plots. The results of this study showed that all methods produced similarly high correlation and concordance coefficients, indicating good accuracy as a method. Limits of agreement analysis indicated that the population level performance of Sun's prediction equations was very similar to the performance of both BIS methods. However, BIS methods in practice have slightly better predictive performance than the single-frequency equations as judged by higher correlation and the limits of agreement from the Bland-Altman analysis. In any case, the authors believe that an accurate evaluation of performance of the methods cannot be done as long as the evaluation is done using Bland-Altman analysis, the commonly accepted technique for this kind of performance comparisons. |