Investigation of chemical modifiers for the direct determination of arsenic in fish oil using high-resolution continuum source graphite furnace atomic absorption spectrometry
Autor: | Bernhard Welz, Javier del Campo Menoyo, Tarcísio S. Almeida, Eduardo Carasek, Daniel L. G. Borges, Jörg Feldmann, Éderson R. Pereira |
---|---|
Rok vydání: | 2015 |
Předmět: |
Analytical chemistry
chemistry.chemical_element Food Contamination 02 engineering and technology Standard solution Mass spectrometry 01 natural sciences Mass Spectrometry Analytical Chemistry Arsenic Fish Oils Industry Graphite Detection limit Aqueous solution Spectrophotometry Atomic 010401 analytical chemistry Temperature 021001 nanoscience & nanotechnology 0104 chemical sciences Dilution chemistry Calibration 0210 nano-technology Graphite furnace atomic absorption Water Pollutants Chemical Palladium |
Zdroj: | Talanta. 150 |
ISSN: | 1873-3573 |
Popis: | High-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS) has been applied for the development of a method for the determination of total As in fish oil samples using direct analysis. The method does not use any sample pretreatment, besides dilution with 1-propanole, in order to decrease the oil viscosity. The stability and sensitivity of As were evaluated using ruthenium and iridium as permanent chemical modifiers and palladium added in solution over the sample. The best results were obtained with ruthenium as the permanent modifier and palladium in solution added to samples and standard solutions. Under these conditions, aqueous standard solutions could be used for calibration for the fish oil samples diluted with 1-propanole. The pyrolysis and atomization temperatures were 1400 °C and 2300 °C, respectively, and the limit of detection and characteristic mass were 30 pg and 43 pg, respectively. Accuracy and precision of the method have been evaluated using microwave-assisted acid digestion of the samples with subsequent determination by HR-CS GF AAS and ICP-MS; the results were in agreement (95% confidence level) with those of the proposed method. |
Databáze: | OpenAIRE |
Externí odkaz: |