Emergency CT head and neck imaging: effects of swimmer’s position on dose and image quality
Autor: | Z. Deak, Robert Stahl, Michael Scherr, Stefan Wirth, Lucas L. Geyer, Fabian G Mueck, Sebastian Roesch, Michael Seidenbusch |
---|---|
Rok vydání: | 2014 |
Předmět: |
Adult
Male Emergency Medical Services medicine.medical_specialty Image quality Posture Computed tomography dose index Radiation Dosage Correlation Humans Medicine Radiology Nuclear Medicine and imaging Swimming Aged Neuroradiology business.industry Ultrasound Dose-Response Relationship Radiation General Medicine Middle Aged Cervical Vertebrae Mann–Whitney U test Female Tomography Radiology Tomography X-Ray Computed business Head Body mass index Neck |
Zdroj: | European Radiology. 24:969-979 |
ISSN: | 1432-1084 0938-7994 |
DOI: | 10.1007/s00330-014-3105-1 |
Popis: | To compare the effects of different arm positions on dose exposure and image quality (IQ) in cervical spine CT after trauma in different patient groups. Patients in standard (STD = 126) and in swimmer’s position (SWIM = 254) were included. Body mass index (BMI subgroup 1 = underweight to subgroup 4 = obese), anterior–posterior diameter (AP), left–right diameter (LR), area of an ellipse (AoE) and angle between the humeral heads (optimal STD 10°) were used as grouping criteria. Computed tomography dose index (CTDI) was documented. Two radiologists rated the IQ at three levels (CV1/2, CV4/5, CV7/T1) using a semi-quantitative scale (0 = not diagnostic, 1 = diagnostic with limitations, 2 = diagnostic without limitations). The Mann–Whitney U test correlations of grouping criteria with dose effects and intra-class correlation (ICC) were calculated. ICC was 0.87. BMI grouping showed the strongest correlation with dose effects: CTDI of optimal STD versus optimal SWIM positioning was 3.17 mGy versus 2.46 mGy (subgroup 1), 5.47 mGy versus 3.97 mGy (subgroup 2), 7.35 mGy versus 5.96 mGy (subgroup 3) and 8.71 mGy versus 8.18 mGy (subgroup 4). Mean IQ at CV7/T1 was 1.65 versus 1.23 (subgroup 1), 1.27 versus 1.46 (subgroup 2), 1.06 versus 1.46 (subgroup 3), 0.79 versus 1.5 (subgroup 4). Patients with a BMI > 20 kg/m2 benefited from both potential dose reduction and improved image quality at the critical cervicothoracic junction when swimmer’s position was used. • BMI is a useful metric for personalized optimization in CT for the c-spine. • Using swimmer’s position, patients can benefit from dose reduction. • In some patients a superior image quality can be achieved with swimmer’s position. • For swimmer’s positioning an angle of more than 10° is optimal. |
Databáze: | OpenAIRE |
Externí odkaz: |